ProChem SSW Steam Sterilization Integrator is a single use Class 5 Integrating Indicator. It integrates three essential criteria for proper steam sterilization: time, temperature, and steam. When processed along with a load it indicates, with a margin of safety, whether the proper steam Rauserv c.c. in Johannesburg, Gauteng, South Africa can supply equipment to suit your application, Medical, Dental or Laboratory. Full service and repairs division available to assist at all times.
Adequate sterilization conditions are reached when the dark bar has completely traveled through the reject window and has entered the accept window.
Use in 115-138˚C gravity, flash or pre-vacuum cycles.
ProChem SSW in the presence of steam will respond to time and temperature.
Pack Size: 250 strips per pack
Expiry: 3 Years
Class: 5 Integrator Indicator
Standards: Complies with ANSI/AAMI/ISO 11140-1:2005, ANSI/AAMI ST60:1996
Temperature Time in Minutes for ProChem SSW to Reach Accept Window 121˚C 24 132˚C 2.2 134˚C 1.5
Temperature Time in Minutes for ProChem SSW to Reach Accept Window.